Details -- Canto 1 (V073000535) -- 2 lasers/8 colors

The Canto 1 is sometimes referred to as a Canto A.  It is equipped with an HP xw4400 workstation using Windows XP and FACSDiva software, Version 6.1.3 (build 2009.05.12.13.29).  Firmware 1.11, CST 1.1.3, PLA 1.1.3.

The Canto 1 has BD’s QC protocol (CST – Cytometer Setup and Tracking) run on it daily to verify CVs and trends in performance. It is kept under BD service contract; so it receives preventative maintenance (PM) every 6 months and is kept to BD’s performance specs.  The machine is cleaned weekly.This cytometer was manufactured in March 2006 and put into service at the CIC in April 2006.

FSC = photodiode with 488/10 filter*

Octagon (Blue Laser, 488nm solid state, 20mW)

 

 

Filter 1

Filter 2

A

PE-Cy7

735LP

780/60

B

PerCp-Cy5.5

655LP

670LP

C

PE

556LP

585/42

D

FITC

502LP

530/30

E

SSC

---

488/10

F

(no detector)

---

---

G

(no detector)

---

---

H

(no detector)

---

---

 

Trigon (Red Laser, 636nm, HeNe, 17mW)

 

 

Filter 1

Filter 2

A

APC-Cy7

735LP

780/60

B

APC

---

660/20

C

(no detector)

---

---

 

*in order to assist a subset of users with their work in macrophage studies, the Canto 1 currently has a BD 0.5 neutral density filter installed on the FSC detector.  This filter elevates the BrightBead CVs slightly.

 

Maintenance:

The CIC flow facility is run by Shawn “Trip” Gill, a Systems Engineer with 15 years of experience working with BD flow cytometers.

Basic reagents (sheath fluid, shutdown solution, FACSClean solution) are all BD products used within BD guidelines.

 

Further technical specs can be found on the BD website:

http://www.bdbiosciences.com/instruments/facscanto/index.jsp